Lyophilized peptide storage for receiving labs

Keep the manufacturer vial lyophilized, closed, dark, and dry at the labeled temperature until the experiment needs solvent.

Last reviewed: 14 October 2026

Answer. Common labeled storage is −20 °C, or −80 °C if the laboratory SOP requires it. Most avoidable damage happens from moisture on a cold cake, or from treating one reconstituted vial as a working bottle.

Before the vial is opened

ControlLaboratory practiceWhy it matters
TemperatureStore at the labeled condition in a controlled freezer, not a frost-choked door shelf.Freeze–thaw of the unopened vial is still a temperature cycle.
MoistureKeep caps closed. Open only after the vial has approached room temperature in a dry box or desiccator.Condensate on a cold cake starts hydrolysis and clumping.
Light and airDark storage. Minimize time with the cap off.Oxidation of Met, Cys, and Trp begins at the headspace.
LocationAssigned freezer ID in the batch file. Quarantine bin until CoA and label match.A verified batch that cannot be found is not a verified batch.
Opened vs unopenedDo not open a vial “to check” the cake as a receiving step.Receiving inspects the closed container, the label, and the documents.

After the laboratory decides to use the batch

  • Keep unused vials lyophilized at the labeled condition.
  • Reconstitute only what the work requires, or split immediately into single-use frozen aliquots.
  • Freeze once per aliquot. Thaw once. Do not refreeze.
  • Working solutions stay on ice or at bench only for the length of the assay.
This page does not assign a universal shelf life after reconstitution, a multi-puncture calendar, or a product-specific solvent volume. See also freeze–thaw and the excursion form.

Short questions

Is a melted shipper automatic rejection?
No. It is an excursion note. Keep the vial closed and decide with the batch file.
Should receiving labs store opened working stocks in the refrigerator for weeks?
Not as a default. Prefer single-use frozen aliquots.