08 — Definitions

Identity vs purity vs peptide content

They are three different numbers. Mixing them up is how stock concentrations go wrong.

Last reviewed: 14 October 2026

Answer. Identity asks “is this the intended molecule?” Purity asks “how much of the measured peak is that species?” Peptide content asks “how much of the weighed solid is peptide rather than counter-ion and water?”

The three measurements

TermTypical methodQuestion it answersWhat it does not tell you
IdentityMass spectrometry (MS)Does the observed mass match the intended sequence?How much related impurity is in the vial
PurityHPLC at a stated wavelengthWhat share of the detected peak area is the main species?Absolute milligrams of peptide in the solid
Peptide content / net peptideElemental analysis, AAA, or a stated content methodWhat fraction of the net weight is peptide?Whether the sequence is correct
Net weight (as-is)Label mass of the lyophilized solidHow much material is in the vial as shipped?Peptide-only mass if salt and water are present

Why the distinction matters in a notebook

Lyophilized peptides often include counter-ions (commonly TFA or acetate) and residual moisture. A vial labeled 10 mg as-is is not automatically 10 mg of peptide. If the CoA reports peptide content separately, use that basis for any mass/volume stock and write the basis in the notebook.

How to read the methods line

  • HPLC purity without a wavelength is incomplete. Different wavelengths change the apparent %.
  • MS identity without a purity method does not quantify related substances.
  • A high purity % and a lower peptide-content % can both be consistent.
  • Do not transfer a % from one lab’s method onto another lab’s chromatogram.

Short questions

If HPLC reports a high percentage, is the vial that percentage peptide by weight?
Not necessarily. That figure is usually peak-area purity, not peptide content.
Which figure should a receiving lab file?
All three that the CoA actually reports, plus the batch number on the vial.
Does a high purity number mean the material is approved for people?
No. Analytics do not change intended use.